← Clinical Trials
Patients Indicated for an ICD Phase not stated Binary Endpoint NCT00836589

Long-term Evaluation of the Linox Family ICD Leads Registry (GALAXY): Complete Statistical Analysis

The GALAXY registry followed the Linox family ICD leads with BIOTRONIK ICDs, with the stated objective of assessing long-term safety and reliability. The primary endpoint was a serious-adverse-event-free rate through 5 years, but the primary result and its confidence interval are not reported in the registry record.

Trial start: 2008-12  ·  Enrollment: 1997  ·  Sponsor: Biotronik, Inc.
Scope of this record

This page provides an independent statistical analysis and educational interpretation of publicly reported results. ClinicalTrials.gov provides the official trial registry record. Numerical trial results on this page are restricted to the results posted in the ClinicalTrials.gov record; calculations added by Clinical Biostats are labelled as such.

1. Trial at a Glance

This page covers the registry results for Long-term Evaluation of the Linox Family ICD Leads Registry (GALAXY) (NCT00836589), a registered phase not stated study. The primary outcome is Percentage of Subjects Who Are Free of Complications Related to the Linox ICD Lead.

1997
Enrollment
Phase not stated
n/a
Arms
In the primary outcome
n/a
Effect
No re-analysis shown
n/a
95% CI
See limitations
FeatureRegistry value
ClinicalTrials.gov identifierNCT00836589
PhasePhase not stated
ConditionPatients Indicated for an ICD
AllocationNot Stated
Design modelNot Stated
MaskingNot Stated
Enrollment1997
Arms in the primary outcomen/a
Lead sponsorBiotronik, Inc.
Results postedYes

2. Clinical Question

Population

Patients indicated for an ICD were enrolled. The primary endpoint’s evaluable population was defined as the total number of unique subjects who completed the 5-year follow-up and/or experienced a primary endpoint adverse event. The number meeting that definition is not reported in the registry record.

Intervention

The registry collected data on ICD therapy using the Linox Lead System implanted with a market-released BIOTRONIK ICD device. The study objective was to characterize long-term lead safety and reliability, including failures that could contribute to loss of pacing or defibrillation therapy.

Comparator

No comparative control group is described. The registry lists a Data Collection Group, but allocation is not stated, and the record does not establish a like-for-like control comparison. The study therefore addresses outcomes observed in the registry cohort rather than a treatment contrast between randomized groups.

Primary question

What percentage of evaluable subjects remained free of complications related to the Linox ICD lead through 5 years? A complication was described as a serious adverse event requiring additional invasive intervention to resolve or specific non-invasive actions. The endpoint was expressed as a serious-adverse-event-free rate. The numerical rate and its 95% confidence interval are not reported in the registry record, so the primary question cannot be answered quantitatively from the posted results.

3. Trial Design

GALAXY is a registry study of the Linox family ICD leads. Its stated objective was to confirm long-term safety and reliability when the Linox Lead System was used with BIOTRONIK ICDs. The registry aimed to characterize lead failures from implant through 5 years, including failures contributing to loss of pacing or defibrillation therapy. The listed intervention is ICD therapy in an ICD lead registry, and the listed group is a Data Collection Group.

The enrollment was 1997. The registry record lists a start date of 2008-12, primary completion in 2016-12, and status as completed. Allocation is not stated. The record does not describe randomized assignment or a control group, so these data do not constitute a randomized comparison. A between-group treatment effect is not defined by the information in the record.

The primary outcome concerns the percentage of subjects free of Linox ICD lead-related complications over 5 years. The endpoint description defines these complications as serious adverse events requiring additional invasive intervention to resolve or specific non-invasive actions. The evaluable population consists of unique subjects completing 5-year follow-up and/or experiencing a primary endpoint adverse event. The number of evaluable subjects is not reported in the registry record.

Secondary outcomes include serious adverse events excluded from the primary outcome, pacing threshold, sensing, and pacing impedance measurements at scheduled follow-ups through 5 years, and electrical parameters by Linox lead model. The registry record lists 10 outcome measures and one analysis. Details of follow-up completeness, measurement schedules, and the statistical plan beyond the endpoint description are not reported in the registry record.

4. Participant Flow and Accounting

The registry reports how participants moved through each stage of the study. All counts below are exactly as posted; missing counts are not inferred from other stages.

Overall Study

StageData Collection Group
STARTED1997
COMPLETED731
NOT COMPLETED1266

Reasons for not completing this period:

ReasonData Collection Group
Lost to Follow-up246
Death429
Withdrawal by Subject or Physician591

Analysis population for the primary outcome, as defined in the registry: The evaluable subject population is the sum of the total unique subjects who completed the 5-year follow-up and/or who experienced a primary endpoint adverse event.

5. Baseline Characteristics

The table reproduces the baseline characteristics posted in the registry, which describe the participants analysed at baseline and help show whether the randomized groups were comparable.

CharacteristicUnitData Collection Group
Participants analysed1997
Age, Customized: Age at Enrollmentyears64.5 (SD 12.6)
Sex: Female, Male: FemaleParticipants537
Sex: Female, Male: MaleParticipants1460
Race/Ethnicity, Customized: Race – WhiteParticipants1298
Race/Ethnicity, Customized: Race – Black or African AmericanParticipants288
Race/Ethnicity, Customized: Race – Hispanic or LatinoParticipants244
Race/Ethnicity, Customized: Race – UnknownParticipants139
Race/Ethnicity, Customized: Race – AsianParticipants21
Race/Ethnicity, Customized: Race – Native Hawaiian or Other Pacific IslanderParticipants4
Race/Ethnicity, Customized: Race – Amerian Indian or Alaska NativeParticipants3
Heightin68.0 (SD 4.0)
Weightlbs194.2 (SD 49.4)

6. Primary Endpoint

The registry lists the primary outcome as Percentage of Subjects Who Are Free of Complications Related to the Linox ICD Lead, measured over 5 years.

ElementRegistry entry
Result typeNumber
Measure of spread95% Confidence Interval
Unitpercentage of subjects
Endpoint classificationBinary Endpoint

The primary endpoint is the percentage of subjects who are free of complications related to the Linox ICD lead through 5 years. The record describes the endpoint as a serious-adverse-event-free rate. A qualifying complication was a serious adverse event requiring additional invasive intervention to resolve or specific non-invasive actions. This definition focuses on complications considered consequential enough to require an intervention or action, rather than every reported event.

The endpoint is intended to summarize lead-related safety over extended follow-up. A complication-free proportion can give a direct account of how many evaluated subjects remained free of the defined event during the specified follow-up period. Its interpretation depends on the endpoint definition, the population included in the denominator, and the handling of follow-up that is incomplete or unavailable.

The registry defines its evaluable population as unique subjects who completed the 5-year follow-up and/or experienced a primary endpoint adverse event. The actual evaluable denominator, observed complication-free rate, and 95% confidence interval are not reported in the registry record. Consequently, the endpoint’s numerical estimate and precision cannot be assessed from the posted results. The endpoint also does not, by itself, provide a comparison with another lead or device strategy.

7. Why No Re-analysis Is Shown

Reported as posted, not recalculated. Not a randomized comparison (allocation: not stated); arms are not a like-for-like control comparison. A result is recalculated only when the posted data support a clean two-arm comparison, so no result is shown beyond what the sponsor posted.

8. Statistical Interpretation

Clinical Biostats interpretation

The registry’s primary outcome is a five-year serious-adverse-event-free rate for complications related to the Linox ICD lead. Although the record provides the endpoint definition and names a 95% confidence interval as its measure of uncertainty, it does not report the primary rate, the interval limits, or the evaluable denominator. The main numerical result therefore cannot be described, and its precision cannot be judged from the registry record.

This absence matters because a percentage alone would describe the observed frequency in the evaluated population, while its confidence interval would convey the uncertainty associated with that estimate. General statistical interpretation: a confidence interval gives a range of values reasonably compatible with the observed data under the analysis assumptions. Its width depends in part on the amount of information available and the number of observed outcomes. Without the reported estimate, denominator, and interval, neither the scale of the observed rate nor the degree of uncertainty can be evaluated.

The record reports adverse-event counts for the Data Collection Group: 215 serious adverse events among 1997 at risk, 31 other adverse events among 1997 at risk, and 429 deaths. These entries are not a substitute for the primary endpoint result. The serious adverse event count is not identified in the record as the number of subjects meeting the primary lead-related complication definition; the primary endpoint specifically concerns lead-related complications and uses its own evaluable population. The record also lists the serious adverse events excluded from the primary outcome as secondary outcomes, making the distinction between primary and other adverse-event categories important.

Serious adverse-event terms listed include infection not due to secondary infection at 1.4%, ICD lead impedance out of range, low impedance, potential insulation break at 1.3%, ICD lead dislodgement occurring at or before 180 days post-implant procedure at 1.2%, and several terms at 0.9%. These are listed term percentages, not the primary complication-free rate. The denominator and whether subjects could contribute to more than one term are not reported in the registry record.

The registry has no stated allocation and does not establish a comparable control group. Its posted data can therefore describe events in the cohort, but they do not estimate a randomized contrast or establish how outcomes would differ under another device or strategy. The listed outcomes support a broad safety and device-performance description, but numerical interpretation is limited by the missing primary estimate and details about the analysis population and missing follow-up.

9. Statistical Methods Explained

The primary measure is a percentage of subjects free of a defined complication through 5 years. In general, a percentage summarizes the share of a specified analysis population meeting an outcome definition. Its meaning depends on who enters the denominator and what qualifies as an event. In GALAXY, the primary event is a lead-related serious adverse event requiring additional invasive intervention to resolve or specific non-invasive actions. The analysis population is described as unique subjects who completed 5-year follow-up and/or experienced a primary endpoint adverse event.

The record names a 95% confidence interval for the primary outcome. General explanation: a confidence interval expresses uncertainty around an estimate under the model and assumptions used for the analysis. A narrower interval indicates greater precision than a wider interval, while neither interval width nor the estimate alone establishes a causal treatment effect. The primary rate and the confidence interval limits are not reported in the registry record, nor is the number of subjects in the defined evaluable population. The estimate and its precision therefore cannot be reviewed.

The endpoint is not described as a time-to-event estimate. It is reported in the record as a percentage of subjects, with the timeframe of 5 years. General explanation: a percentage at a follow-up horizon and a time-to-event analysis answer related but distinct questions. A time-to-event method uses the timing of events and can account for differing observation times under its assumptions. A simple proportion instead depends on the defined denominator and how subjects with incomplete follow-up are handled. The registry record does not report further details about the primary analysis method or handling of incomplete observation.

The secondary outcomes include pacing threshold, sensing, and pacing impedance measurements at scheduled follow-ups through 5 years, as well as individual electrical parameters by Linox lead model. These are measurement outcomes rather than the primary complication-free percentage. General explanation: repeated measurements describe device or lead characteristics over follow-up, and their interpretation depends on the timing of measurements, the number of observations at each visit, and how missing measurements are handled. Those details and the numerical measurement results are not reported in the registry record.

Adverse-event summaries require attention to the unit being counted. The record reports serious adverse events, other adverse events, and deaths for the Data Collection Group, with 1997 at risk for both adverse-event categories. It does not state whether the reported counts represent unique subjects, event episodes, or a particular analysis population. Nor does it give the relationship between those counts and the primary endpoint’s evaluable population. The listed serious-event terms are reported as percentages, but their denominators and rules for multiple terms per subject are not reported in the registry record. General explanation: event counts and subject counts can differ when a person experiences more than one event, so the counting unit is essential for interpreting a summary.

The registry record lists one analysis, but a detailed statistical analysis plan is not reported in the registry record. In particular, the record does not specify adjustment methods, subgroup analysis procedures, or approaches to missing data. No method should be presumed from the endpoint name or the fact that a confidence interval is specified. The registry also lists multiple secondary outcomes. General explanation: when many outcomes are examined, the chance of observing a seemingly notable result by chance can increase, depending on the testing strategy. Whether and how multiplicity was addressed is not reported in the registry record.

Planning illustrations based on an observed effect are not the trial’s observed power. The listed planning assumptions are a two-sided significance level of 0.05, equal allocation between two arms, and a normal approximation, with effects to detect expressed as fractions of the observed effect. These are planning assumptions, not evidence that this registry had randomized arms or used that design. Allocation is not stated, and the record does not describe a like-for-like control comparison. Such illustrations should not be read as results from the enrolled cohort or as a substitute for the missing primary estimate.

10. Exact Estimates for Each Arm

No two-arm comparison is shown for this trial, but each arm's result can still be summarized with exact interval estimates computed from the posted counts. The Clopper-Pearson interval is exact and conservative; the Wilson interval is a common closer-to-nominal alternative; the Bayesian credible interval uses a uniform Beta(1, 1) prior. For small groups the three can differ noticeably.

ArmParticipantsEventsObserved %Clopper-Pearson 95% CIWilson 95% CIBayesian 95% credible interval
Data Collection Group77770690.9%88.6% to 92.8%88.6% to 92.7%88.6% to 92.7%

R code and output

arm    <- c("Data Collection Group")
n      <- c(777)
events <- c(706)
for (i in seq_along(n)) {
  ci <- binom.test(events[i], n[i])$conf.int          # Clopper-Pearson exact interval
  cr <- qbeta(c(0.025, 0.975), 1 + events[i], 1 + n[i] - events[i])   # Bayesian, uniform Beta(1,1) prior
  wi <- (function(x, n) { p <- x / n; z <- qnorm(0.975); d <- 1 + z^2 / n
    c((p + z^2 / (2 * n) - z * sqrt(p * (1 - p) / n + z^2 / (4 * n^2))) / d, (p + z^2 / (2 * n) + z * sqrt(p * (1 - p) / n + z^2 / (4 * n^2))) / d) })(events[i], n[i])
  cat(sprintf("ARM|%s|%d|%d|%.1f|%.1f|%.1f|%.1f|%.1f|%.1f|%.1f\n", arm[i], n[i], events[i], 100 * events[i] / n[i],
              100 * ci[1], 100 * ci[2], 100 * wi[1], 100 * wi[2], 100 * cr[1], 100 * cr[2]))
}
ARM  Data Collection Group  777  706  90.9  88.6  92.8  88.6  92.7  88.6  92.7

11. Multiplicity, Missing Data and Analysis Populations

The registry lists a primary endpoint and multiple secondary outcomes, including adverse events, pacing threshold, sensing, and pacing impedance measurements, with some measures described by lead model. The record does not state how the family of outcomes was handled statistically. Whether formal multiplicity adjustment was planned or performed is not reported in the registry record. General explanation: testing or interpreting many outcomes can increase the opportunity for chance findings, so readers should distinguish a prespecified primary question from secondary or exploratory measurements. No conclusion about a multiplicity procedure can be drawn from the listed outcomes alone.

The primary endpoint uses an evaluable population defined as unique subjects who completed 5-year follow-up and/or experienced a primary endpoint adverse event. The number of subjects in that population and the number who completed follow-up are not reported in the registry record. The record does not describe how missing endpoint status, incomplete follow-up, withdrawal, or unavailable measurements were handled. It also does not report the extent or pattern of missing data. These omissions prevent assessment of whether the reported population, if a numerical primary result were available, represents all enrolled subjects or a selected subset.

The registry gives an adverse-event summary with 1997 at risk for serious and other adverse events, but the relation of those risk sets to the primary endpoint population is not stated. A risk set for an adverse-event summary should not automatically be treated as the denominator for the primary complication-free rate. The primary result and its denominator are not reported in the registry record.

A detailed statistical analysis plan, including any rules for multiplicity or missing data, is not reported in the registry record. General explanation: prespecified rules help clarify which analyses were intended and how incomplete observations affect estimates. Without those rules and the associated numerical information, the robustness of the endpoint analysis cannot be evaluated.

12. Safety Results

The registry reports serious adverse events as affected participants divided by participants at risk.

ArmParticipants with serious adverse eventsParticipants at riskPercentage
Data Collection Group215199710.8%

Deaths

ArmDeaths (participants affected of at risk)
Data Collection Group429 of 1997

Other (non-serious) adverse events

Participants with at least one non-serious event at or above the registry's 0% reporting threshold.

ArmParticipants affected of at risk
Data Collection Group31 of 1997

Most frequent serious adverse event terms

Event termOrgan systemData Collection Group
Infection not due to secondary infectionInfections and infestations27 of 1997
ICD lead impedance out of range, low impedance, potential insulation breakCardiac disorders25 of 1997
ICD lead dislodgement (occuring <= 180 days post-implant procedure)Cardiac disorders24 of 1997
ICD lead impedance out of range, high impedance, potential conductor fractureCardiac disorders18 of 1997
RA lead dislodgementCardiac disorders18 of 1997
Other ICD lead related: lead noise or oversensingCardiac disorders17 of 1997
Infection due to secondary infectionInfections and infestations14 of 1997
ICD lead high pacing threshold, intermittent capture, no lead captureCardiac disorders12 of 1997
HematomaSurgical and medical procedures7 of 1997
ICD lead dislodgement (occuring > 180 days post-implant procedure)Cardiac disorders6 of 1997

Most frequent non-serious adverse event terms

Event termOrgan systemData Collection Group
HematomaSurgical and medical procedures12 of 1997
RA lead dislodgementCardiac disorders11 of 1997
ICD lead high pacing threshold, intermittent capture, no lead captureCardiac disorders2 of 1997
ICD lead impedance out of range, high impedance, potential conductor fractureCardiac disorders2 of 1997
RA lead oversensing or noiseCardiac disorders2 of 1997
RA lead impedance out of range, high impedance, potential conductor fractureCardiac disorders1 of 1997
Infection not due to secondary infectionInfections and infestations1 of 1997

Serious events with the highest arm-level rate: infection not due to secondary infection (1.4%), icd lead impedance out of range, low impedance, potential insulation break (1.3%), icd lead dislodgement (occuring <= 180 days post-implant procedure) (1.2%), icd lead impedance out of range, high impedance, potential conductor fracture (0.9%), ra lead dislodgement (0.9%).

Do not overextend the safety comparison. Serious-adverse-event counts are safety observations, not efficacy outcomes, and should not be combined with the primary result into a single benefit-risk number.

The registry reports 215 serious adverse events among 1997 at risk, 31 other adverse events among 1997 at risk, and 429 deaths for the Data Collection Group. These are posted safety counts and should be kept distinct from the primary endpoint: the primary outcome concerns lead-related complications, uses a defined evaluable population, and is expressed as a serious-adverse-event-free rate. The primary rate and its denominator are not reported in the registry record.

The record lists serious-event terms with percentages, including infection not due to secondary infection at 1.4%, low impedance or potential insulation break at 1.3%, early ICD lead dislodgement at 1.2%, and several other terms at 0.9%. Those term percentages are not interchangeable with the primary endpoint. Their denominators and whether a subject could contribute to more than one term are not reported in the registry record. General explanation: a term-level percentage describes the reported frequency for that term under its counting rules; without its denominator and counting unit, it should not be interpreted as a subject-level risk.

The registry record does not explain how deaths relate to lead complications, follow-up completion, or the primary evaluable population. These counts alone do not establish the cause of events or a comparative safety difference.

13. What the Result Does, and Does Not, Mean

The GALAXY registry was designed to characterize long-term safety and reliability of the Linox Lead System used with BIOTRONIK ICDs, including lead failures that could contribute to loss of pacing or defibrillation therapy. The registry identifies a five-year primary endpoint based on the percentage of evaluable subjects free of specified lead-related complications. That objective and endpoint definition describe the question of interest, but the primary numerical result is not reported in the registry record.

Without the primary rate, its 95% confidence interval, and the evaluable denominator, the posted results do not permit a quantitative assessment of the observed complication-free proportion or its precision. The adverse-event counts and term percentages provide safety information, but they cannot be substituted for the endpoint because they may refer to different event definitions, populations, or counting units. The record does not provide enough detail to resolve those relationships.

The registry is not described as randomized, and allocation is not stated. The record does not establish a like-for-like control group. Accordingly, the cohort data do not estimate a randomized comparison between the Linox lead system and another device, nor do they establish what outcomes would have occurred under an alternative strategy. Descriptive event data and causal comparative evidence answer different statistical questions.

The record also lists electrical measurements and additional adverse-event outcomes. These could characterize device performance and safety if their numerical results and analysis details were available, but those values are not reported in the registry record. Details of missing follow-up, measurement availability, multiplicity, and the statistical plan are also absent. General explanation: conclusions about long-term percentages depend not only on observed events, but on who was followed, how outcomes were defined, and how incomplete observations were treated.

The appropriate reading is therefore bounded: the registry record establishes the intended five-year safety question and reports selected adverse-event summaries, but it does not report the primary endpoint estimate needed to answer that question quantitatively. The posted information does not establish comparative effectiveness or prove the presence or absence of a treatment effect.

14. Why This Trial Matters Statistically

Long-term device registries raise statistical questions that differ from those in a short, controlled comparison. The GALAXY registry targeted complications and lead failures from implant through 5 years, including failures that could contribute to loss of pacing or defibrillation therapy. Such follow-up makes the definition of the analysis population and the handling of incomplete observation central to interpretation. The record defines an evaluable population, but does not report its size or how missing follow-up was handled.

The primary endpoint combines a clinical event definition with a subject-level percentage. Readers need to know which adverse events qualify, which subjects enter the denominator, and how uncertainty is quantified. GALAXY specifies a serious-adverse-event-free rate and a 95% confidence interval, but does not report the rate or interval limits. This illustrates an important statistical distinction: a well-defined endpoint is not itself a result, and an uncertainty measure cannot be evaluated when the estimate is absent.

The registry also lists multiple forms of evidence, including adverse-event summaries and electrical measurements such as pacing threshold, sensing, and pacing impedance. General explanation: complementary outcomes can describe different aspects of safety and device performance, but they are not interchangeable. Their denominators, follow-up availability, and analysis methods matter. Where those details are not reported, conclusions must remain limited to what the record explicitly states.

Finally, the registry does not describe a randomized comparison or a control group. Its central statistical contribution, if the primary result were reported, would be descriptive evidence about the observed cohort rather than a randomized estimate of a difference between lead strategies. The absent primary result and analysis details limit even that descriptive interpretation. Recognizing these limits helps keep the registry’s stated aim separate from what its posted numerical results can establish.

15. Limitations

The most important limitation for interpreting the primary question is that the registry record does not report the primary serious-adverse-event-free rate, its 95% confidence interval, or the number of subjects in the evaluable population. The endpoint is defined, but its magnitude and precision cannot be assessed. The record also does not report how many enrolled subjects completed 5-year follow-up.

The evaluable population includes unique subjects who completed follow-up and/or experienced a primary endpoint adverse event. Details explaining how this rule was implemented, how subjects with incomplete follow-up were treated, and how other missing data were handled are not reported in the registry record. This limits assessment of the denominator and potential effects of incomplete observation.

The safety summaries list 215 serious adverse events, 31 other adverse events, and 429 deaths, with 1997 at risk for each adverse-event category. The record does not clarify the event-counting unit, the relation of these summaries to the primary endpoint population, or how deaths were handled in relation to lead outcomes. The denominators for the listed serious-event term percentages are also not reported in the registry record.

Allocation is not stated, and no like-for-like control group is established. Therefore, the record does not provide a randomized comparison. A detailed statistical analysis plan, including procedures for multiplicity, missing data, and analysis of the secondary electrical measurements, is not reported in the registry record. These omissions constrain both quantitative interpretation and assessment of how robust any analysis may be.

16. Related Tutorials

17. Related Statistical Calculators

18. Sources

Continue through the Clinical Biostats statistical pathway

Use this trial's endpoint as a starting point for deeper study of effect measures, confidence intervals, and sample-size planning.

19. Record Summary

The completed GALAXY registry enrolled 1997 patients indicated for an ICD to evaluate the long-term safety and reliability of the Linox Lead System used with BIOTRONIK ICDs. The registry record lists a start date of 2008-12 and primary completion in 2016-12. Its primary endpoint was the percentage of subjects free of Linox ICD lead-related complications through 5 years, expressed as a serious-adverse-event-free rate.

The endpoint’s evaluable population was defined as unique subjects who completed 5-year follow-up and/or experienced a primary endpoint adverse event. The primary estimate, 95% confidence interval, and evaluable denominator are not reported in the registry record. The record lists adverse-event counts of 215 serious events, 31 other events, and 429 deaths, and includes secondary outcomes involving adverse events and electrical measurements. Allocation is not stated, and the record does not establish a control comparison. Further details on the statistical plan, missing data, and multiplicity are not reported in the registry record.